Neural Signal Amplifier. Part Numbers: 4208, 5703, 5749, 5748, 5747. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP)
FDA Recall Z-2763-2016 · Posted 2016-09-08
Advertisement
Recalling FirmBlackrock Microsystems, LLC
Recall StatusTerminated
Product CodeGWL
Date Initiated2016-08-10
Date Terminated2017-08-21
Product Quantity81 total amplifiers
Reason for Recall
Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type CF Applied Parts.
Firm Action
Blackrock Microsystems sent an Urgent Medical Device letter dated August 12, 2016 via email. The letter identified the affected product, problem and actions to be taken. Customers were instructed not use the amplifier in application requiring a type CF applied part; perform a physical count of the inventory of amplifiers and patient cables with the part numbers listed. For questions regarding this Correction call 801-582- 5533, ext. 240.
Root Cause
Labeling False and Misleading
Distribution Pattern
Worldwide Distribution - US Nationwide and Canada,