Terminated

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

FDA Recall Z-2763-2019 · Posted None

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeCGR
Date Initiated2018-09-18
Date Terminated2024-02-21
Product QuantityAccess Cortisol Reagent Assay

Reason for Recall

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

Firm Action

An Urgent Medical Device Recall Letter was sent during the week of 9/17/2018 to all affected customers via mail and or email. Customers were instructed to discontinue use and discard remaining product. Customers were also instructed to consider a review of the Cortisol results generated with the affected lot number at the discretion of the Medical Director.

Root Cause

Process design

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑