SureSight Obturator Introducer Localization Systems (ILS) for ATEC in MRI. ILS-0914-20-OB (9G Introducer Localization Set (OB Model) for ATEC0914-20MR Handpiece)
FDA Recall Z-2765-2016 · Posted 2016-09-08
Advertisement
Recalling FirmHologic, Inc
Recall StatusTerminated
Product CodeGDF
Date Initiated2016-08-02
Date Terminated2017-03-30
Product Quantity1440 units
Reason for Recall
Tip Breakage at the glue joint between the obturator body and tip
Firm Action
Hologic sent an Important Product Recall letter to customers in the United States through Certified letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately examine their inventory and cease use of all affected recalled product. Customers were asked to complete the enclosed Recall Response Card indicating whether you have any of the recalled product on hand, as well as the quantity of product that you have in stock.
Root Cause
Process control
Distribution Pattern
US Nationwide Distribution
Official Source