Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

FDA Recall Z-2771-2026 · Posted 2026-07-22

Recall StatusOpen, Classified
Product CodeLLZ
Date Initiated2026-06-19
Product Quantity68 units

Reason for Recall

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

Firm Action

Boston Scientific initially notified their consignees via teleconference on 6/19/2026 and followed with a notice distributed on the same date via email. The notice explained the issue, potential harms, and requested the following: "Required User Actions " Follow the instructions provided in the TheraSphere 360 Ordering Quick Guide. " Remove and reselect vials if inputs into the Activity calculator (Lung Shunt Fraction, Anticipated Residual Waste, and/or Target Volume) are changed after any vials were added to the cart. " Verify that the vial information displayed in the cart and treatment documentation match the final calculation inputs. " Review order information in the attachment to ensure treatment and documentation are as intended. Contact your TheraSphere Consultant to address any impacted treatments or with any questions."

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑