Recall StatusOpen, Classified
Date Initiated2026-06-16
Product Quantity3,087 units
Reason for Recall
Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
Firm Action
Olympus notified consignees on about 06/16/2026 via letter. Consignees were informed that they many continue to use their UHI-4 until the offset adjustment is performed. However, as noted in the Instructions for Use (IFU), prepare a spare UHI-4 unit as a backup to ensure uninterrupted operation can be completed without complication in the case of a malfunction. Actions to take included to carefully read the notification letter and ensure all personnel are completely knowledgeable on the content of the notification, an Olympus field service representative will contact you to schedule a mutually convenient time to perform the offset adjustment, acknowledge receipt of the notification through the Olympus recall web portal, and forward the notification to other users who may have the affected products if you have further distributed it.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, El Salvador, Germany, Honduras, Hong Kong, India, Ireland, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Paraguay, Peru, Russia, Singapore, Sri Lanka, Taiwan, Thailand, United Kingdom, Uruguay, Vietnam.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.