Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.

FDA Recall Z-2778-2015 · Posted 2015-09-17

Recall StatusTerminated
Product CodeFDT
Date Initiated2015-06-30
Date Terminated2017-05-25
Product Quantity256 units

Reason for Recall

The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.

Firm Action

FujiFilm sent a Voluntary Field Correction/Verification Form dated June 30, 2015 to their affected customers.

Root Cause

Device Design

Distribution Pattern

Distributed in the US and Canada.