Biosense Webster MobiCath Bi-Directional Guiding Sheath, Model D140010 (small curve) and D140011 (large curve). The MobiCath Bi-Directional Guiding Sheath is a flexible tipped percutaneous catheter introducer (or guiding sheath) designed for gaining access to the coronary systems. The device features adjustable tip geometry through use of a rotating handle to deflect the catheter. This catheter is assembled with a hemostasis valve and a sideport infusion line. The kit includes a flexible catheter dilator to facilitate bi-directional guiding sheath passage.

FDA Recall Z-2782-2017 · Posted None

Recalling FirmGreatbatch Medical
Recall StatusTerminated
Product CodeDYB
Date Initiated2017-02-15
Date Terminated2019-05-09
Product Quantity11,360

Reason for Recall

One side of the MobiCath Bi-Directional Guiding Sheath pouch may not be sealed.

Firm Action

Consignee was sent a Greatbatch "Urgent Medical Device Recall" letter dated February 15, 2017 via UPS and email . The letter was addressed to Biosense Webster, Inc. The letter described the product involved in the recall, Reason for Recall, Risk to Health,& Instructions to Biosense Webster. Advised consignee to quarantine and return the devices, complete and return the field Recall Response Form to Stericycle, and to communicate the recall to their customers and end users. For questions contact your local Greatbatch Medical representative, Monday through Friday via telephone at 1-763-951-8184 from 8:00 AM (CST) to 5:00 PM (CST), or e-mail us at [email protected]

Root Cause

Process control

Distribution Pattern

CA