Siemens YSIO - solid state x-ray imager Solid state x-ray imager (flat panel/digital imager), image-intensified fluoroscopic x-ray system

FDA Recall Z-2787-2011 · Posted 2011-07-11

Recall StatusTerminated
Product CodeMQB
Date Initiated2011-04-28
Date Terminated2014-04-10
Product Quantity141

Reason for Recall

There is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is possible that parts of the table may collide with extremities and cause injuries.

Firm Action

Siemens sent a Safety Advisory Notice to all affected customers. The Safety Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to place the Safety Notice in the System's instruction for use. If product was distributed further, customers were to forward the safety notice to the new owner and inform the firm of the new owner. Customers were instructed to complete the Acknowledgement of Receipt Form and return to the firm.

Root Cause

Other

Distribution Pattern

Nationwide Distribution including Washington, DC and the following States: AR, CA, CO, FL, ID, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.