Terminated

Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.

FDA Recall Z-2789-2016 · Posted 2016-09-13

Recall StatusTerminated
Product CodeMBH
Date Initiated2016-08-01
Date Terminated2017-06-08
Product Quantity2,039 units

Reason for Recall

Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.

Firm Action

Stryker sent an "Urgent Medical Device Recall Notification" Letter and Recall Business Reply Form dated August 1, 2016 to their affected customers via email and also via UPS on August 1, 2016. Customers are asked to quarantine the product subject to this recall. Branches/Agencies are asked to return all affected products available at their location to: Stryker C/O Stericycle, Attn: RA 2016-098 - Event 8296. Hospital/Branches: Customers are asked to complete and sign the Business Reply Form and fax a copy within 5 business days to: 877-272-6141 or email to Stericycle [email protected]. Any questions, customers can call (201) 831-6693.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide and Internationally

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →