Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds
FDA Recall Z-2792-2026 · Posted 2026-07-30
Advertisement
Recalling FirmMedical Action Industries, Inc. 306
Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2026-05-13
Product Quantity657
Reason for Recall
Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within blood vessels.
Firm Action
On 5/13/2026, Owens & Minor emailed recall notices to customers who were asked to do the following:
Identify, segregate, and quarantine all impacted product. Other components within the convenience kits can still be used once kit is properly labeled with the warning label below.
Add warning labels to affected Medical Action kits to ensure all users are aware of the actions to remove and discard Manifold component as well as actions identified below and in the enclosed Medline recall notice. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Labels state in part: Prior to use, remove & discard the Medline Namic Angiographic (The Manifold).
Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling.
Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness.
Complete and return the response form via email to [email protected]
Report any quality problems or adverse events associated with the products listed in this notification to [email protected]
Root Cause
Material/Component Contamination
Distribution Pattern
US: AZ, CO
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.