SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System

FDA Recall Z-2793-2026 · Posted 2026-07-23

Recalling FirmAlphatec Spine, Inc.
Recall StatusOpen, Classified
Product CodeGWF
Date Initiated2026-05-13
Product Quantity3

Reason for Recall

Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.

Firm Action

On 5/13/2026, recall notices were emailed to customers who were advised of the following: 1) Recalling firm has provided replacement SafeOp 3 Systems. 2) Abstain from using affected devices and return to recalling firm. 3) Share this notice within your organization or any organization where the affected device(s) has(have) been transferred. 4) Fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions, as described in this notification. If you have any questions/concerns or require assistance with the recall, please e-mail [email protected] or call 1-800-922-1356.

Root Cause

Software Design Change

Distribution Pattern

US Nationwide distribution in the states of TX, MO, IN.