Dynamic Locking Screw System (DLS) 3.7mm & Dynamic Locking Screw (DLS) 5.0mm; intended for use in long bone fractures, the fixation of osteopenic bone, the fixation of osteotomies, and for the fixation of non-unions and malunions.
FDA Recall Z-2794-2015 · Posted 2015-09-18
Advertisement
Recalling FirmSynthes (USA) Products LLC
Recall StatusTerminated
Product CodeHWC
Date Initiated2015-08-11
Date Terminated2016-08-18
Product Quantity40038
Reason for Recall
DLS pin breakage during planned implant removal, after uneventful and successful healing of the fracture. This notice to additional consignees identified after recall Z-0390-2014, related to complaints of breakage located at the bottom of the screws and under the locking screw head.
Firm Action
Synthes sent an Urgent Notices of Medical Device Recall letter, dated August 11, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers were instructed to return affected devices and response forms.
For further questions please call (610) 719-6500.
Root Cause
Labeling False and Misleading
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.