3.5mm Non Locking Cortical Screw, in the PediLoc Tibial Plate System Part number 00-0903-2510, 3.5mm x 10mm Non Locking Cortical Screw. Used in pediatric orthopedics.

FDA Recall Z-2796-2015 · Posted 2015-09-18

Recalling FirmOrthoPediatrics Corp
Recall StatusTerminated
Product CodeHRS
Date Initiated2015-07-28
Date Terminated2016-03-17
Product Quantity5

Reason for Recall

The 00-0903-2510, 3.5mm x 10mm Non Locking Cortical Screw, Lot # 150192-J package may contain a different screw. The screw in the package is a 00-0903-2526, 3.5mm x 26mm, and is etched correctly.

Firm Action

OrthoPediatrics has initiated a Voluntary RECALL of Part # 00-0903-2510, 3.5mm Non Locking Cortical Screw, Lot # 150192-J, initially by contacting customers via telephone, and then by e-mailing notification letters on 7/28/2015. Customers returned the 5 mislabled screws, which were repacked with correct labeling. All recall activities were complete as of 8/14/2015.

Root Cause

Packaging process control

Distribution Pattern

Domestic: TN, IA Foreign: England VA/DOD: None