Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Neuro Pack, Pack Number AMS12195C; 2) Neuro Pack, Pack Number AMS12195D; 3) Neuro Pack, Pack Number AMS12307; 4) Neuro Spine Pack, Pack Number AMS16313; 5) Lami Kukkar Pack, Pack Number AMS17958; 6) Neuro Pack, Pack Number AMS6452B; 7) Neuro Lami Pack, Pack Number AMS9427C.

FDA Recall Z-2801-2026 · Posted 2026-07-24

Recall StatusOpen, Classified
Product CodeOJG
Date Initiated2026-06-08
Product Quantity927 units

Reason for Recall

The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.

Firm Action

Aligned Medical Solutions (AMS) issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 6/8/2026 via email. The notice explained the issue, potential risk, and requested the following actions: identify and quarantine affected product, affix the sticker which reads: "RECALL NOTICE Medicom announced the recall of Ritmed Neurological Sponges contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall." " At the time the kit is opened for use any Medicom-Ritmed Neurological Sponges should be identified and set aside. " The recalled Medicom-Ritmed Neurological Sponges should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Complete the recall reply form and return it via email to: [email protected] even if you have no affected product on hand. Those further distributing the product were directed to notify their distribution partners. For questions contact [email protected].

Root Cause

Material/Component Contamination

Distribution Pattern

US Nationwide distribution in the states of AZ, CA, CT, GA, IL, MA, MD, NY, OH, PA, TN, TX, VA, WA.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑