Arrow TheraCath Epidural Catheter - Product Usage: The Arrow Epidural Catheter permits access to the epidural space.
FDA Recall Z-2803-2020 · Posted None
Advertisement
Recalling FirmArrow International Inc
Recall StatusTerminated
Product CodeCAZ
Date Initiated2020-07-06
Date Terminated2021-11-05
Reason for Recall
The lidstock states the incorrect expiration date for the product. This issue could result in use of a device that is expired which could potentially lead to an increased risk of infection or other complications. Sterility, biocompatibility, safety, or efficacy of the kits and their components are not assured beyond the correct expiration date.
Firm Action
Arrow sent a customer notification letter dated July 6, 2020. This letter informed customers of the affected product, asked them to check and remove inventory, and complete the response form.
Root Cause
Incorrect or no expiration date
Distribution Pattern
World wide distribution