NextStep Retrograde Replacement Hub Set - Product Usage: is indicated for use in the replacement of a NextStep Retrograde hub connection assembly that has been damaged.

FDA Recall Z-2804-2020 · Posted None

Recall StatusTerminated
Product CodeNFK
Date Initiated2020-07-06
Date Terminated2021-11-05

Reason for Recall

The lidstock states the incorrect expiration date for the product. This issue could result in use of a device that is expired which could potentially lead to an increased risk of infection or other complications. Sterility, biocompatibility, safety, or efficacy of the kits and their components are not assured beyond the correct expiration date.

Firm Action

Arrow sent a customer notification letter dated July 6, 2020. This letter informed customers of the affected product, asked them to check and remove inventory, and complete the response form.

Root Cause

Incorrect or no expiration date

Distribution Pattern

World wide distribution