Open, Classified

L-Cath Midline S/L Catheter 20ga (3F) 1mm x 20cm, REF: 384620

FDA Recall Z-2804-2026 · Posted 2026-07-24

Recall StatusOpen, Classified
Product CodeLJS
Date Initiated2026-06-15
Product Quantity20

Reason for Recall

Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.

Firm Action

On 6/15/2026, recall notices were emailed to customers who were asked to do the following: 1) Provide this recall notice to all people who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2) Segregate, quarantine, and return affected devices to recalling firm. 3) Complete and return the response form via email [email protected] For additional information on this letter, please email [email protected]

Root Cause

Device Design

Distribution Pattern

US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU