M5 Corpus Power Wheelchair REF,120032

FDA Recall Z-2805-2026 · Posted 2026-07-24

Recalling FirmPermobil Inc
Recall StatusOpen, Classified
Product CodeITI
Date Initiated2025-06-19
Product Quantity715 wheelchairs

Reason for Recall

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

Firm Action

On June 10, 2025 Permobil issued a technical bulletin (recall notice) via Email to affected consignees. Permobil informed consignees of the following: 1. Cellular connection is required to receive and install the firmware update. 2. It is important to not use the wheelchair when the firmware update is in process (as indicated on the display of the input device). 3. Specific Remote Update Instructions are available if needed in addition to the on-screen information. 4. A confirmation of the completed update will be shown on the joystick display when the update successfully completes. 5. If you have any questions or concerns, please contact the Permobil Customer Experience team at (800) 736-0925. 1.

Root Cause

Process design

Distribution Pattern

Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.