Fujifilm Processor EPX 2500 High Definition Endoscopy; (Sales Brochure). For use in endoscopic procedures. Used to observe, record and capture images.

FDA Recall Z-2806-2015 · Posted 2015-09-22

Recall StatusTerminated
Product CodeFET
Date Initiated2015-06-30
Date Terminated2017-05-25
Product Quantitymultiple

Reason for Recall

The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.

Firm Action

Fujifilm Medical Systems USA, Inc. initiated this recall by sending a Voluntary Field Correction Letter and Tracking/Verification Form dated June 30, 2013 to their affected customers.

Root Cause

Device Design

Distribution Pattern

Nationwide US and Canada