Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.
FDA Recall Z-2811-2015 · Posted 2015-09-23
Advertisement
Recalling FirmSynthes (USA) Products LLC
Recall StatusTerminated
Product CodeGXN
Date Initiated2015-08-31
Date Terminated2016-08-18
Product Quantity105882
Reason for Recall
The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
Firm Action
DePuy Synthes sent an URGENT:FIELD SAFETY NOTIFICATION / MEDICAL DEVICE LABELING CORRECTION letter, dated August 31, 2015 to affected customers. The letter identified affected product, problem and actions to be taken.
Customers were instructed to:
"Update and review the package insert (GP1345-D).
" Discard outdated package insert noted in the table above.
" Forward this Field Safety Notification to anyone in your facility that needs to be informed, especially personnel that conduct MR testing.
" If the package insert has been forwarded to another facility, contact that facility.
" Complete the attached Verification Section (page 3 of this notification). Please include your name, title, address, telephone number and signature in the spaces provided. Send a copy of the completed Verification Section by, Fax: 877-884-7821 or Scan/email: [email protected]
" If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page three (3) of this notification.
" Maintain a copy of this notice.
For questions call 610-719-5450.
Root Cause
Labeling False and Misleading
Distribution Pattern
US Nationwide and internationally, including Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement