Terminated

61 cm (24") Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, MicroClave Clear, 1.2 Micron Filter, 3 Gang 4-Way NanoClave Stopcock (Glow, Blue, Red Rings), Rotating Luer, 2 Ext.

FDA Recall Z-2811-2018 · Posted None

Recall StatusTerminated
Product CodeFMG
Date Initiated2018-03-01
Date Terminated2024-01-26
Product Quantity100 pouches

Reason for Recall

ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.

Firm Action

ICU Medical notified customers of the recall on about 03/01/2018 via URGENT: Medical Device Recall Notification. The notification informed customers that this was an update and expansion to previously initiated recall. Instructions included to inspect inventory for affected devices, arrange for the return of affected devices, and complete and return the Urgent Medical Device Recall Notification Response Form to [email protected].

Root Cause

Other

Distribution Pattern

US distribution to AZ, MA, NC, OH, TX, VA, WA. Worldwide distribution to Belgium, France, Germany.