PAPR (Powered Air-Purifying Respirator) - Model# WLPAPR20W1
FDA Recall Z-2815-2020 · Posted None
Advertisement
Recalling FirmWhirlpool Corporation
Recall StatusTerminated
Product CodeNZJ
Date Initiated2020-07-02
Date Terminated2021-06-25
Product Quantity528 units
Reason for Recall
The rivets that secure the elastic headband to the face shield have the potential to be mis-assembled. When this condition occurs, it can result in a
headband that can become loose from face shield allowing the face shield to shield on the user.
Firm Action
Customers were contacted via email/telephone beginning 6/18/20.
Rework was conducted in-house at the customer facilities.
Root Cause
Manufacturing material removal
Distribution Pattern
The products were distributed to the following US states: MI.