Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381

FDA Recall Z-2816-2020 · Posted None

Recall StatusTerminated
Product CodeHRS
Date Initiated2020-07-21
Date Terminated2022-02-18
Product Quantity191 units

Reason for Recall

May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw

Firm Action

DePuy Synthes issued Urgent Medical Device Removal letter dated July 21, 2020 stating reason for recall, health risk and action to take: Please take the following actions: 1. Immediately examine your inventory to determine if you have product subject to this action on hand and quarantine the subject product. 2. Complete the attached Verification Section (page 3 of this letter). Please include your facility name and address, account number, name of person completing the form, title, email address, telephone number and signature in the spaces provided. If you have product to be returned: " Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. " Return the completed Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132 " Send a copy of the completed Verification Section (page 3 of this letter) to DePuy Synthes by: Scan/email: [email protected] " Return any subject product within 10 business days. If you have no product to be returned: " Return the Verification Section (page 3 of this letter) to DePuy Synthes by: Scan/email: [email protected] 3. Forward this notice to anyone in your facility that needs to be informed. 4. If any of the subject product has been forwarded to another facility, contact that facility to arrange return. 5. Keep a copy of this notice visibly posted for awareness until the product subject to this action has been returned. While processing your returns, please maintain a copy of this notice with the product subject to this action and keep a copy for your records. Questions, please call 610-719-5450 or contact your DePuy Synthes Sales Consultant

Root Cause

Process control

Distribution Pattern

US Nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑