ASP-1020 LipoTower with Aspiration Pump. Medical use.

FDA Recall Z-2821-2015 · Posted 2015-09-30

Recall StatusTerminated
Product CodeMUU
Date Initiated2015-06-12
Date Terminated2016-05-23
Product Quantity3 LipoTower with Aspiration Pump

Reason for Recall

There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.

Firm Action

The firm has issued FDA recall letters to all consignees. The letter have been sent out by UPS to expedite the delivery.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution.