ASP-1021 LipoTower with Aspiration Pump and Tumescent Pump. Medical use.
FDA Recall Z-2822-2015 · Posted 2015-09-30
Advertisement
Recalling FirmMicroAire Surgical Instruments, LLC
Recall StatusTerminated
Product CodeMUU
Date Initiated2015-06-12
Date Terminated2016-05-23
Product Quantity168, ASP-1021 LipoTower with Aspiration Pump and Tumescent Pump
Reason for Recall
There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.
Firm Action
The firm has issued FDA recall letters to all consignees. The letter have been sent out by UPS to expedite the delivery.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.