NICO Shepherd' s Hook - Greenberg, sterile. An accessory to NICO s BrainPath device Part Number: NN-8024
FDA Recall Z-2822-2018 · Posted None
Advertisement
Recalling FirmNico Corp.
Recall StatusTerminated
Product CodeGZT
Date Initiated2018-06-19
Date Terminated2018-11-09
Product Quantity204 devices
Reason for Recall
Mislabeled hook pouch included a transport/storage label rather than the product label containing key product information such as part number, lot number, and expiration date, etc
Firm Action
On June 19, 2018 , NICO issued letter by mail and/or email letters to consignees that have received hooks from the affected lot. The letter provides instructions to inspect all of its current stock of NN-8024, Greenberg Hooks for the proper pouch label. if an incorrect label is discovered, the hook should be
removed from stock and returned to NICO.
Questions or need additional information, please contact [email protected] or 317-660-7118 ext. 104.
Root Cause
Packaging process control
Distribution Pattern
US Nationwide distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.