P/N: 9734856, SCREWDRIVER, 9734856, SOLERA STD MAST; P/N: 9734857, SCREWDRIVER, 9734857, SOLERA RDN MAST; P/N: 9734279, SCREWDRIVER 9734279 SOLERA STANDARD; and P/N: 9734373, SCREWDRIVER 9734373 SOLERA REDUCTION. The Medtronic Navigated Solera Screwdrivers are reusable devices that are designed for use in spine surgery to assist in the placement of screws during the surgery, using a Medtronic computer-assisted surgery system.

FDA Recall Z-2824-2015 · Posted 2015-09-30

Recall StatusTerminated
Product CodeHAW
Date Initiated2015-09-14
Date Terminated2015-12-03
Product Quantity2346

Reason for Recall

Specific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.

Firm Action

Medtronic sent an Urgent Medical Device Correction Notification letter dated September 14, 2015, to all affected consignees via FedEx. Revised Instructions for Use were provided to add additional warnings related to the careful inspection of the instruments and the importance of knowledge of operating procedures, patient selection, and product information. Consignees with questions were instructed to to call the local Medtronic Navigation Technical Services Representative or call 1-800-995-9709.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution - US (nationwide) and Internationally to Canada; United Arab Emirates; Australia; Belgium; Switzerland; China; Germany; Denmark; Spain; Finland; France; Great Britain; Greece; India; Israel; Italy; Japan; Netherlands; New Zealand; Poland; Portugal; Sweden; Singapore; Thailand; Slovakia; Turkey; and South Africa.