Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Arrow Kits with Medtronic Covidien 0.9% Sodium Chloride Flush Syringes

FDA Recall Z-2825-2015 · Posted 2015-09-30

Recall StatusTerminated
Product CodeFOZ
Date Initiated2015-08-27
Date Terminated2018-06-18
Product Quantity73714

Reason for Recall

Manufacturer of 0.9% Sodium Chloride Flush Syringes recalled due to compromised sterility of the outer barrel, Arrow purchased recalled items that were packaged with certain products.

Firm Action

An urgent recall notification letter, dated August 27, 2015, was sent to direct consignees to inform them of the issue and provided instructions for removing the syringes from the kit and returning product along with the response form.

Root Cause

Process control

Distribution Pattern

Nationwide and internationally.