Recall StatusTerminated
Date Initiated2015-08-03
Date Terminated2016-01-20
Product Quantity72 units
Reason for Recall
Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.
Firm Action
Ferno began calling their customers on 8/3/2015. The firm informed their customers of the following:
To provide optimal Customer Experience and minimal disruption to your service, the iNX field action involves the following steps:
" Exchange of your current iNX with a new iNX;
" Participation in a brief training session; and
" Inspection of your Ferno Model 175 antler and rail fastening system as
requested by the Dept. of Transportation.
As you await delivery of your new iNX, it is important you ensure proper function of the fastening system and iNX telescoping load frame.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.