FPS 35mm x 3.5 mm non-locking screw

FDA Recall Z-2835-2016 · Posted None

Recalling FirmOrtho Solutions Inc
Recall StatusTerminated
Product CodeHWC
Date Initiated2016-07-21
Date Terminated2016-10-12
Product Quantity15

Reason for Recall

Mislabeling of a single batch of FPS 3.5 x 35 mm Locking Screw as being Non-Locking Screws.

Firm Action

Ortho Solutions sent a "Field Safety Notice" letter dated July 21, 2016 to their affected customers. The letter identify the Solutions Business Development Manager will contact the customer to arrange for the isolation and removal of the unused product from their inventory. Replacement product will be provided. The customer was advised to passed on the "Field Safety Notice" to all who need to be aware within their organization and to any organization to which their have further distributed or transferred the affected screw. Upon removal of the affected device, customers are asked to please sign and return the "Field Safety Corrective Action Declaration Form", by which to confirm they have received the "Field Safety Notice" and that they declare that the required recall has been completed at their hospital. For further questions, call (201) 258-7305.

Root Cause

Labeling mix-ups

Distribution Pattern

US Distribution