The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
FDA Recall Z-2835-2018 · Posted None
Advertisement
Recalling FirmIntuitive Surgical, Inc.
Recall StatusOpen, Classified
Product CodeNAY
Date Initiated2018-05-17
Product Quantity1
Reason for Recall
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Firm Action
Intuitive Surgical notified customers via a "Field Safety Notice
Urgent Medical Device Correction ISIFA2018-06-C
da Vinci Xi Miscalibrated Endoscope Controller" letter. The letter provided: Reason for Field Action, Risk to Health, Affected Products (description, part number, serial number), Action taken by Intuitive Surgical, Actions to be taken by the Customer/ User, contact information for Further Information & Support, and an Acknowledgement Form.
Root Cause
Use error
Distribution Pattern
U.S.
Official Source