Terminated

"***3i Incise Custom-Made Label. QP157 Approved 07/09/10*** 3i inc se Custom-made Dental Framework***Biomet 3i 4555 Riverside Drive Palm Beach Gardens, FL 33410 800-342-5454 www.biomet3i.com Please consult instructions for use and recommendation***For Dental Lab use ONLY Non-sterile/For single Use***" Indicated for use in restorative procedure to affix the final prosthesis to the abutment. Coping and frameworks provide the base structure for final porcelain application as the final step of creating a restoration. The devices are intended to be used to create final restoration for dental reconstruction with either a prepared tooth or dental implant abutment.

FDA Recall Z-2838-2011 · Posted 2011-07-16

Recalling FirmBiomet 3i, LLC
Recall StatusTerminated
Product CodeNSP
Date Initiated2010-12-01
Date Terminated2011-07-18
Product Quantity16

Reason for Recall

On December 1, 2010 Biomet 3i recalled their product Incise Zirconia Coping (CBZR0102). The Coping contains a manufacturing defect. Coping is oversized. All affected direct accounts were notified by letter on December 14, 2010.

Firm Action

Biomet 3i sent an "MEDICAL DEVICE RECALL" letter dated December 14, 2010 to all affected customers. The letter identifies the product, problem, and actions to be taken. The letter instructs customers to complete and fax the Product Recall Response form to 561-514-6316. In addition, a shipping label is enclosed with the letter for returning the affected product. Customers with questions and concerns should contact Biomet 3i recall coordinator at 561-776-6906.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution--USA (nationwide) and the country of Canada.