EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter
FDA Recall Z-2854-2016 · Posted None
Recalling FirmEKOS Corporation
Recall StatusTerminated
Product CodeKRA
Date Initiated2016-08-09
Date Terminated2016-10-05
Product Quantity10 units
Reason for Recall
The recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant parameters, which may result in a small blister in the epoxy over transducers in the treatment zone. It is not known at the time the recall was initiated whether the non-compliant test parameters, or presence of the blisters, adversely affect operation of the device.
Firm Action
All affected consignees were contacted by e-mail on August 9, 2016, and asked to sequester implicated devices.
Root Cause
Storage
Distribution Pattern
Three international consignees - Germany, New Zealand and Taiwan