Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
FDA Recall Z-2854-2026 · Posted 2026-07-31
Advertisement
Recalling FirmCopan WASP
Recall StatusOpen, Classified
Product CodeJQP
Date Initiated2026-06-16
Product Quantity1
Reason for Recall
Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.
Firm Action
On 06/16/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was provided to customers informing them that, A software bug has been identified in the WebApp software which, under specific
conditions, may prevent the result suggested by PhenoMATRIX from being overwritten
by the result selected by the user. As a consequence, the result transmitted by WASPLab
to the LIS may be the result suggested by PhenoMATRIX, rather than the result selected
by the user.
Customers were informed and instructed:
In date 2026/06/15 we released a WebApp software version including the bugfix of this
issue. As a consequence, you are required to schedule with your distributor the software
update to correct the issue and, in the meanwhile, to check the results sent to LIS in
case you changed the result suggested by PhenoMATRIX.
Complete the attached Customer Response Form and return to the Copan WASP
Srl contact noted on the form so that Copan WASP Srl may acknowledge your receipt of this notification.
For questions or further assistance email [email protected]
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the state of Arizona.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.