Terminated

PHS Spinal Tray 24G Gertie Marx, Catalogue No. 899379. Intended to provide licensed medical professionals with a kit containing drugs and other medical devices that are used to support the administration of general, local anesthetics.

FDA Recall Z-2868-2016 · Posted None

Recall StatusTerminated
Product CodeCAZ
Date Initiated2016-08-24
Date Terminated2016-12-19
Product Quantity320 packages

Reason for Recall

Integra LifeSciences has recently identified that one lot of Spinal Tray 24 G Gertie Marx Catalogue # 899379 was shipped containing the 24 G Gertie Marx needle without an introducer.

Firm Action

Consignees notified on 8/24/16 via an Urgent - Field Corrective Action letter e-mail. The letter identified the affected device and the reason for the recall. Customers were asked to review their inventory of the Spinal Tray 24 G Gertie Marx to confirm the quantity received. The attached form should be completed and returned by e-mail or fax as indicated. Once the form is received, Customer Service will place an order for the customer to receive replacement devices. Questions should be directed to Customer Service at 1-800-241-2210.

Root Cause

Employee error

Distribution Pattern

Distributed to CA.