Connecting Bolt, Model 14-442093
FDA Recall Z-2875-2018 · Posted None
Advertisement
Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeLXH
Date Initiated2018-06-18
Date Terminated2020-03-26
Product Quantity247 units
Reason for Recall
Product was manufactured from the incorrect material, which shows a potential decrease in the fatigue strength of the connecting bolt that can lead to fracture.
Firm Action
Zimmer Biomet issued URGENT MEDICAL DEVICE RECALL notices to customers via courier service.
Risk Manager were advised to take the following actions:
1. Review the notification and ensure that affected personnel are aware of the contents.
2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.
3. Complete and return the Certificate of Acknowledgement form via email to [email protected]. This form must be returned even if you do not have affected products at your facility.
4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation.
5. If you have further questions or concerns after reviewing the notice, call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Distributors are encouraged to do the following:
Review th notification and ensure that affected team members are aware of the contents.
2. Immediately locate and quarantine affected product in inventory.
3. Immediately return all affected product from your distributorship and from affected hospitals within your territory.
a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days.
b. Request a Return Authorization Number through Fast SMS or by emailing [email protected]
c. For each return, send a copy of Attachment 1 to [email protected].
d. Include a hardcopy of Attachment 1 in each carton of your return shipment for
Root Cause
Employee error
Distribution Pattern
AR, AZ, CA, FL, GA, IN, KY, LA, MA, MI, MN, MO, MT, NJ, NY, OH, SC, TX, UT & WI
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement