WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.
FDA Recall Z-2878-2018 · Posted None
Advertisement
Recalling FirmBiosense Webster, Inc.
Recall StatusTerminated
Product CodeNGV
Date Initiated2018-05-07
Date Terminated2020-11-14
Product Quantity25 units
Reason for Recall
When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.
Firm Action
Coherex Medical notified customers via "URGENT FIELD SAFETY NOTICE
Coherex WaveCrest Left Atrial Appendage (LAA) Occlusion System (32 mm)
Catalog No: WCR1523, Lot Number: All" letter dated May 7, 2018.
The letter provided an overview, and important information about the product description and codes, lots subject to removal, why customers were being contacted, What actions are required/instructions, available assistance/contact information, and Reply Form.
Root Cause
Component design/selection
Distribution Pattern
No U.S. Distribution. International Distribution only.