Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Pediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous access to the central circulation of pediatric patients. The ARROWg+ard antimicrobial surface is intended to help provide protection against catheterrelated infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for long-term use. The use of the 4, 5, and 5.5 Fr. antimicrobial surface catheters in infants weighing less than 2.0, 2.5, and 3.0 kilograms respectively, have not been evaluated. Therefore, the benefits of these catheters in infants weighing less than the indicated weights should be weighed against any possible risks.

FDA Recall Z-2886-2018 · Posted None

Recall StatusTerminated
Product CodeFOZ
Date Initiated2018-07-03
Date Terminated2020-04-29
Product Quantity166

Reason for Recall

One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm catheter. No patient injuries have been reported related to this issue.

Firm Action

On July 3, 2018, the firm sent out Urgent Medical Device Recall letters to customers. The letter communicated the labeling error. Affected consignees are instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot number listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter.

Root Cause

Employee error

Distribution Pattern

Distributors in MO and UT. Medical facilities in AL, CA, IL, MA, NH, NM, NY, TX, VA, and WI.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑