Match Point System The Match Point System is intended to by uses as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-image scans.
FDA Recall Z-2886-2020 · Posted None
Advertisement
Recalling FirmMaterialise N.V.
Recall StatusTerminated
Product CodeKWS
Date Initiated2020-07-17
Date Terminated2020-12-23
Product Quantity128
Reason for Recall
The procedure side indicated in the top header of the Shoulder Case Planning Report incorrectly indicated the procedure side as "Left" for surgical plans for a "right" procedure side.
Firm Action
The firm, materialise, sent Recall letters dated 17 July 2020 informing customers of the issue. This letter informs the customer that Materialise has corrected the Shoulder Case Planning Report in the SurgiCase online system.
If you have any questions, contact Global Quality and Regulatory Manager Medical, at 734-259-7017 or email: [email protected].
Root Cause
Software Design Change
Distribution Pattern
Nationwide distribution to: US including PR and International to UK and Australia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.