Recall StatusOpen, Classified
Date Initiated2026-06-08
Product Quantity43 units
Reason for Recall
Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.
Firm Action
On 06/08/2026 an "URGENT: Medical Device Recall Notification" was sent via First Class Mail and Email and/or Facsimile to customer informing them that as a result of receiving a customer complaint/report, it was confirmed that antibiotic disk cartridges labeled as HardyDisk AST Cefiderocol (disk code FDC30) were found to contain sporadic disks labeled with Piperacillin/Tazobactam disk code (TZP110).
Customers are instructed to:
1. Discard any remaining stock.
2. Response to this notification is requested within 5 business days of receipt of this letter, either through mail or email. To respond by mail, complete, sign, and send the enclosed inventory reconciliation form (FORM-2023C) via USPS First Class Mail. To respond by email, scan and send the completed inventory reconciliation form (FORM-2023C) to
[email protected].
3. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to invoice upon request.
For Questions - contact Technical Services Department at (800) 266-2222, option 2 or via email at
[email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
U.S. Nationwide distribution in the states of AZ, CA, FL, ID, IN, MA, MS, NJ, NV, NY, OH, PA, and TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source