BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
FDA Recall Z-2889-2020 · Posted None
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeFMI
Date Initiated2020-07-28
Date Terminated2021-04-23
Product Quantity125,200 units
Reason for Recall
Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.
Firm Action
BD Sent a customer letter dated 28 July 2020. this letter requests customers to review their inventory for effected lots and Destroy all product subject to the recall. The letter requests customers to return the included response form and notify BD of any adverse events associated to this product.
Root Cause
Equipment maintenance
Distribution Pattern
US and Canada