Terminated

Aquilion One: Aquilion TSX-301B, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning.

FDA Recall Z-2896-2011 · Posted 2011-07-29

Recall StatusTerminated
Product CodeJAK
Date Initiated2011-01-07
Date Terminated2012-08-17
Product Quantity5

Reason for Recall

Toshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may be broken due to failure of the optical conversion connector, resulting in a break of communication between units.

Firm Action

Toshiba America Medical Systems, Inc. sent an "Urgent Medical Device Correction" letter dated January 4, 2011 to all affected customers. The letter provides the customers with an explanation of the problem and the firms plan to install preventive measures and provided them with a Customer Reply form to send Toshiba. A representative will contact the customers for an appointment to install the measures to their CT system. For question on this recall call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.

Root Cause

Software design

Distribution Pattern

(USA) Nationwide Distribution including the states of WI, MO, TN, IA, AL, AZ, TX, and GA.