Siemens ADVIA Centaur Calibrator U, For in vitro diagnostic use in calibrating the following assays using ADVIA Centaur"' systems: Myoglobin cTnl Cat No. ADVIA Centaur Calibrator U (2 Pack) - 03684480 (SMN 1 0309996)

FDA Recall Z-2899-2016 · Posted None

Recall StatusTerminated
Product CodeDDR
Date Initiated2016-08-08
Date Terminated2018-10-03
Product Quantity2,248 kits

Reason for Recall

Failure of analytical sensitivity with Calibrator U kit lots ending in 63 and 64 and a drift from the internal standardization

Firm Action

Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction (UMDC) to customers in the US on August 8, 2016, to be delivered to customers on August 9, 2016. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on August 8, 2016. The communications notify customers of the failure of analytical sensitivity with Calibrator U kit lots ending in 63 and 64 and a drift from the internal standardization. The communications also inform customers that analytical sensitivity and alignment to the internal standardization will be restored with the release of Calibrator U kit lots ending in 65. Customers may continue to use Calibrator U lots CU63 and CU64 with any indate reagent kit lot, current QC ranges and current Master Curve Material lot 45719. Revised QC targets and ranges are located on the Bio-Rad website at QCnet.com for use with CU65 and higher. For further questions, please call (508) 668-5000.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - Nationwide Distribution including the countries of: Argentina Australia Brazil Canada Chile China Ecuador India Japan Korea Mexico Malaysia Peru Philippines Saudi Arabia Singapore Taiwan Vietnam South Africa