Recall StatusOpen, Classified
Date Initiated2026-05-21
Product Quantity109 units
Reason for Recall
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Firm Action
On 5/21/26, the firm began notifying consignees in the US via letter.
Consignees were informed that Werfen will be proactively shipping replacement PAKs to each facility.
Customers were instructed to check inventory for affected product. They should discontinue use of the affected lots and destroy any inventory on hand. Destruction should be documented via the response form provided to the customer and returned to the recalling firm.
Customers should forward the notification to all affected locations at each facility.
On May 28, 2026, the field action was expanded to include affected customers in China.
On June 23, 2026, the field action was expanded to include additional affected lots in Italy and Germany.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.