Integra Straight Ventricular Catheter F8, 25 cm, Sterile EO, Rx only, packaged in a Tyvek pouch and secondary packaged in to Tyvek pouch, sealed in a cardboard box. Silicone elastomer Ventricular Catheter for use with hydrocephalus valves
FDA Recall Z-2905-2016 · Posted None
Advertisement
Recalling FirmIntegra LifeSciences Corp.
Recall StatusTerminated
Product CodeJXG
Date Initiated2016-08-23
Date Terminated2017-03-29
Product Quantity10
Reason for Recall
During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product.
Firm Action
Integra sent an "Urgent Voluntary Medical Device Recall" Letter and Recall Acknowledgement and Return Form dated August 23, 2016 to their affected customer. The letter identified the product, the problem, and the action to be taken by the customer. The customer was advised if they have distributed the product to identify those customers and forward a copy of the enclosed Integra Medical Device Voluntary Recall Notification to those customers. The customer was also asked to review their inventory for the affected product and complete the attached form indicating whether or not they have any of the affected product in their inventory. Customers with questions were instructed to contact Customer Service at 1-800-654-2873.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution to KS only