Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
FDA Recall Z-2906-2026 · Posted 2026-08-06
Advertisement
Reason for Recall
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Firm Action
On June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person.
Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.
Root Cause
Incorrect or no expiration date
Distribution Pattern
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.