Recall StatusTerminated
Date Initiated2018-08-03
Date Terminated2024-09-23
Product Quantity185
Reason for Recall
Order Status Remains Approved When Should Indicate Complete.
Firm Action
On August 2018 the firm sent letter to all their customer with the following instructions; Recommended User Action:
A user with appropriate pharmacy rights can unlock the record. This blanks out the details requiring the user to re-enter the administration details previously entered.
To avoid this locking issue, close the MAR window before navigating to the CWS. This document contains important information for the continued safe and proper use of your equipment.
Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed.
Advise the appropriate personnel working with this product the content of this letter.
Elekta Corrective Actions:
The issue is resolved in MOSAIQ Release 2.65 Beta 02 and in release 2.64 SP10.
In order to meet regulatory requirements, you are required to complete this the Acknowledgement Form and return it to Elekta immediately upon receipt but no later than 30 days.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, WA, WI, WV, and WY
Australia, Canada, Ireland, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, and United Kingdom
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.