Terminated

Aruba Wireless Network Access Points, Model 65 utilized by the Acuity Central Monitoring System. Welch Allyn Protocol, Inc. Intended for use by clinicians for the central monitoring of neonatal, pediatric and adult patients in health care facilities. The system connects to a network of patient monitors to record and analyze physiological data being acquired by the monitors.

FDA Recall Z-2918-2011 · Posted 2011-08-03

Recall StatusTerminated
Product CodeDSI
Date Initiated2011-05-16
Date Terminated2011-08-12
Product Quantity12 units were affected

Reason for Recall

Aruba Wireless Network Access Points, Model Number 65, experienced a higher than expected incidence of dropouts due to external RF interference.

Firm Action

Welch Allyn Protocol, Inc. sent a letter dated May 16, 2011 to affected customer. The letter stated there were concerns over drop-outs in the system recommended the system at this hospital to revert back to the Symbol wireless infrastructure. For questions call Welch Allyn at 315-685-4100.

Root Cause

Software design

Distribution Pattern

Worldwide distribution: (USA) Nationwide distribution including the state of California; and the countries of Singapore; Philippines, France, Italy, Malaysia, Poland, UK, England, UAE, Germany, Spain, Ireland, Portugal, New Zealand, Hong Kong, Thailand, Australia, Canada, and the Netherlands.