Recall StatusTerminated
Date Initiated2020-08-14
Date Terminated2023-05-01
Product Quantity412 units
Reason for Recall
This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in specific situations which could pose a potential risk.
Firm Action
On 8/14/20 the firm notified their customers via letter of the field action, with the following instructions:
How can you help to avoid the potential risk of this issue?
You may continue to use your system ensuring to follow all instructions in the Operator Manual. Do not leave the patient unattended at any time while the patient is positioned on the patient bed.
Siemens Medical Solutions USA, Inc.
2501 N. Barrington Road
Hoffman Estates, IL 60192 USA
In the event of an Emergency-Stop, shutdown or power loss the patient should be manually retracted from the gantry using the patient bed pallet handle and release button.
In the event of touchpad engagement, confirm there are no system errors reported on the PPM and refer to the instructions in the Operator Manual to address proper patient positioning before continuing.
If unintended detector motion is observed, downward pressure from the detector is reported by the patient at any time or the PPM presents a system error (as shown in the picture below), then manually retract the patient from the gantry using the patient bed handle and release button, stop use of the system and contact your customer service representative. Errors or problems that result in patient re-scan or re-dose should be reported to your local Siemens Healthiness representative.
Distribution Pattern
US Nationwide distribution including in the states of Alabama, Alaska, Arizona, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Nebraska, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, West Virginia and Wisconsin.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.