Open, Classified

Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16

FDA Recall Z-2934-2026 · Posted 2026-08-11

Recall StatusOpen, Classified
Product CodeKRO
Date Initiated2026-07-06
Product Quantity1 US; 9 OUS

Reason for Recall

The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.

Firm Action

Foreign manufacturer, Waldemar Link GmbH & Co KG, notified importer LinkBio Corp. on 7/6/2026 via email. LinkBio then notified the single US consignee by email/Docusign on 7/7/2026. Customer was instructed to return the affected product to LinkBio.

Root Cause

Process change control

Distribution Pattern

Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.